IDEXX VetLab® UA Analyser

Easy, efficient and consistent

The IDEXX VetLab® UA™ Analyser is simple to operate, improves work flow and produces consistent, automatically printed results.

 

The IDEXX VetLab UA Analyser connects to the IDEXX VetLab® Station, a laboratory information management system that collects all results into an electronic medical record and prints them on an integrated report for easy interpretation.

 

IDEXX UA™ Strips are designed and validated for veterinary use only, so you don’t need to wonder if your canine, feline or equine results are valid. Strips can be used manually or with the IDEXX VetLab UA Analyser.

IDEXX VetLab UA Analyser

Easy and efficient—Results you can count on in about 80 seconds

 

Standardised operation and automation—The IDEXX VetLab® UA™ Analyser provides consistent results through automation; human error is all but eliminated.

One-button operation—Use a pipette to flood the strip with the urine sample, place it in the analyser, press a button and walk away.

Efficiency—Automatically reads and prints results, saving you time and reducing the chance for mistakes.

Printout—Results are included in the IDEXX VetLab® integrated report, offering you a tool you can use with your clients as you explain the results and recommendations.

Speed—The IDEXX VetLab UA Analyser provides results in approximately 80 seconds.

Reflectance photometer reads IDEXX UA Strips

 

IDEXX VetLab UA Analyzer TechnologyThe IDEXX UA test strip is placed on a sliding tray that moves the strip into the analyser and under the reading head. The analyser reads the reference pad first, and then reads each of the test pads on the strip.

Note: Only IDEXX UA Strips can be used with the IDEXX VetLab UA Analyser.

 

Measuring Principle

The analyser contains light-emitting diodes (LEDs) that emit light at various wavelengths.

Reading is done electro-optically:

  • The LED emits light of a defined wavelength onto the surface of the test pad at an optimum angle.
  • The light hitting the test zone is reflected more or less intensely depending on the colour produced on the test pad, and is picked up by the detector, a phototransistor positioned directly above the test zone.
  • The phototransistor sends an analogue electrical signal to an A/D converter, which changes it to digital form.
  • The microprocessor then converts this digital reading to a relative reflectance value by referring it to a calibration standard.
  • Finally, the analyser compares the reflectance value with the defined range limits (reflectance values that are programmed into the analyser for each parameter) and outputs a semiquantitative result.

 

Each test pad on the IDEXX UA Strip is read photometrically, producing results in approximately 80 seconds.

 

IDEXX VetLab UA Analyser Specifications
  
DimensionsWidth: approximately 6 in (150 mm)
Depth: approximately 11.5 in (290 mm)
Height: approximately 3.7 in (94 mm)
Weight: approximately 1.7 lb (0.8 kg)
Power SupplyExternal power adapter, Model SA 125A-0735U-S (Sino-American)
Input: 100—240 V AC, 50/60 Hz, 800 mA
Output: 7.5 V DC, 3000 mA
ConsumptionOperating: maximum 15 W
Standby: 1.3 W
System DescriptionType: reflectance photometer
Light source: 6 LEDs
Wavelengths: 565 nm (green) 3x
610 nm (orange) 3x
Reader head: 1 head with 6 LEDs
Measuring cycle: approximately 80 seconds
Maximum throughput: approximately 45 strips/hour
Printer: thermal printer
Display: liquid crystal display, two lines of 24 characters each
Memory: 100 samples
Date/time: integrated clock
Operating ConditionsTemperature: 59º–90ºF (15º–32ºC)
Storage: -4º–158ºF (-20º–70ºC)
Optimum Operating ConditionsTemperature: 68º–78ºF (20º–26ºC)
Relative humidity: 30%–60%

Test Parameters

 

The IDEXX VetLab® UA™ Analyser results print with all your IDEXX VetLab® in-house analyser results on one comprehensive report, so you can see a total picture of your patient's health.

 

The UA analyser prints the results in the following gradation of concentration:

 

IDEXX VetLab UA Analyzer Parameter Chart

 

Manually enter the following information on the patient report printout (figure 1) or on the integrated diagnostic report generated by the IDEXX VetLab Station:

  • Collection method (Coll. Meth.)
  • Volume (Vol.)
  • Colour
  • Clarity
  • Specific gravity (SG)
  • Sediment analysis (Sed.)

 

 


Figure 1

Results in approximately 80 seconds

 

The IDEXX VetLab® UA™ Analyser is easy to use!

 

1.
Apply urine to the strip.
2.
Blot the strip.
3.
Place it in the analyser.

 

4.
Press the button, walk away.
 
 

 

That’s it!
Results are ready in approximately 80 seconds and integrated into your complete IDEXX VetLab patient report.

Frequently asked questions for the IDEXX VetLab® UA Analyser

 

What is a complete urinalysis (UA)?

A complete UA includes physical, chemical and microscopic examinations:

  1. Observation of urine colour and clarity
  2. Urine specific gravity measured by veterinary refractometer
  3. Measurement of urine chemical properties (i.e. pH, bilirubin, glucose, occult blood and ketones)
  4. Microscopic examination of urine sediment

 

When is complete urinalysis indicated?

Primary indications for complete UA include:

  • Any time you suspect a urinary tract infection or urinary tract disease
  • Whenever a CBC and clinical chemistry are indicated
  • Monitoring disease progression or response to therapy
  • Senior Preventative Healthcare Programs: Complete UA is an element of minimum laboratory database in aged (> 7 years old) cats and dogs.1
  • Debilitated or ill patients: Kidneys can be damaged by a variety of non-renal diseases such as immune-mediated disorders, neoplasia, sepsis, tick-borne diseases, viral infections, heartworm infection, severe inflammatory disorders (e.g. dental disease, inflammatory bowel disease, pyoderma) and metabolic disorders (e.g. hyperadrenocorticism, hyperthyroidism).
  • Hypoalbuminaemia: Complete UA is an important tool for the diagnosis and localisation of protein-losing glomerulopathies and for differentiating this from other causes of decreased serum albumin concentration.
  • Clinical suspicion of renal disease: Complete UA is warranted when clinical and laboratory information suggest renal disease, such as azotemia, polyuria/polydipsia, pollakiuria, or radiographic and ultrasonographic kidney abnormalities.
  • To determine the cause for inappropriate elimination or urinary tract-related behavioral disorders

 

Reference
1.
AAHA Senior Care Guidelines for Dogs and Cats [see Journal of the American Animal Hospital Association (2005) 41:81–91] and AAFP and Academy of Feline Medicine, Panel Report of Feline Senior Care [see Journal of Feline Medicine and Surgery (2005) 7:3–32].

 

Does the method of urine collection matter?

Yes, collection method and urine storage container can influence results or affect ability to compare subsequent or follow-up UA results.

  • It is recommended for an individual patient that all complete UAs be performed on urine collected in a similar manner.
  • Owner-improvised urine containers should be discouraged.
  • Sterile containers provided by a hospital are recommended with urine intended for bacteriologic culture and sensitivity.

 

Can any method of urine collection be used with IDEXX UA Strips?

Yes. Midstream-voided, transurethral catheterisation and cystocentesis methods can be used. However, to enable accurate comparison of follow-up testing, it is recommended that the same collection method be used for all UAs on a single patient.

 

What animals can be tested with IDEXX UA Strips?

IDEXX UA Strips are validated for canine, feline and equine urine.

 

What parameters can be tested with the IDEXX UA Strips?

The IDEXX UA Strips measure:

pH
Protein
Ketones
Bilirubin
Leukocytes
Glucose
Urobilinogen
Blood or haemoglobin

 

Why should leukocyte results be verified with microscopy?

Studies in cats show that leukocyte esterase tests are nonspecific for the detection of pyuria, which means that most urine will be positive for leukocytes when none are present. In both dogs and cats, leukocyte esterase tests are insensitive for the detection of pyuria, which means that urine can be negative for leukocytes when leukocytes are present.

 

Why should urine dipsticks not be used to read specific gravity in veterinary samples?

Urine specific gravity, as offered on other urine dipsticks, was designed for use in human medicine only. For veterinary samples, which have a significantly broader range of pH, the test is unreliable. As noted in Fundamentals of Veterinary Clinical Pathology by Stockham and Scott (pg 304), “the reagent strip method is not recommended for estimation of urine solute concentration or specific gravity in urine of domestic mammals.” Furthermore, urine dipstick specific gravity pads only measure up to 1.025 and are inadequate for measuring the urine concentrating abilities of dogs and cats, which are typically greater than 1.040. The refractometer is more precise and has a more appropriate testing range.

 

Why should urine dipsticks not be used to read nitrite in veterinary samples?

Due to the high number of false negatives in small animals, nitrites have a very limited value in veterinary samples.

 

Can I dip the whole strip in the urine sample or should I apply urine to each pad with a syringe?

The recommended procedure is to use a pipette to flood the IDEXX UA Strip with the urine sample, making sure that all the test pads are moistened.

Because of the matrix design of IDEXX UA Strips, runoff of reagent from one test to its neighboring tests is very unlikely. Immediately after flooding the IDEXX UA Strip with urine, the long edge and back side of the strips should be blotted on an absorbent paper to remove excess urine.

 

Can I apply urine even after it has been spun down for sediment exam?

While this is not the recommended procedure, the supernatant of a recently centrifuged urine sample may be used. It is important to note in the medical record that urine supernatant was used for sample analysis.

 

Can urine be used if it was collected in a non-sterile container or from the floor?

Non-sterile urine can be used, but a special note should be made about the container. Depending on the previous use of the container or condition of the surface, urine pH as well as other tests can potentially be affected. Non-sterile urine should never be used to assess bacteria on the microscopic exam.

 

Can I read the IDEXX UA Strips manually after the UA analyser has evaluated them?

Yes, a strip can be read manually after the analyser has evaluated it. But it should be done within two minutes from when urine was applied to the strip. After two minutes, the strip results are not diagnostically significant.

Note: The analyser takes approximately 80 seconds to read a strip.

 

Can I use the same strip twice?

The IDEXX UA Strips can be used only once.

 

What urine strips can be used with the IDEXX VetLab UA Analyser?

The IDEXX VetLab UA Analyser has specific design requirements for the strips that it can read. Therefore, only IDEXX UA Strips should be used with the analyser.

 

How often does the analyser need calibration?

Calibrate the IDEXX VetLab UA Analyser once a week using IDEXX VetLab® UA™ Calibration Strips. The calibration procedure takes less than two minutes. Please refer to the Quick Reference Guide or Operator’s Manual for the calibration procedure.

 

Why are there no reference ranges for the urinalysis results?

A urinalysis includes a semi-quantitative measurement of both chemical contents and formed elements such as cells and casts. During different states of hydration and normal kidney function, these components will have variable concentrations in the urine making it impossible to identify one set of reference intervals. Interpret results in conjunction with the urine concentration (urine specific gravity).

 

If the urine is blood-tinged or opaque will that affect results?

IDEXX UA Strips have a colour compensation pad. The IDEXX VetLab UA Analyser reads the compensation pad first and automatically adjusts for the variability in the urine colour. This will help to minimise any artifacts or changes from discoloured urine. Moderate to severe alterations in urine colour should be noted in the record and the results should be analysed with these changes in mind.

 
Related Products