IDEXX > Companion Animal > In-Clinic Analysers > VetTest Chemistry Analyser > Specifications
 
 

IDEXX VetTest® Specifications

 
Species   Canine, feline, equine, bovine, avian, ferret, goat, lizard, llama, monkey, mouse, pig, rabbit, rat, sheep, snake and tortoise

Units   U.S., French or S.I. units.

Sample Size   The VetTest requires a minimum of 40 μL of sample to run one test and an additional 10 μL for each additional test.

Result Timing   The VetTest analyser takes approximately six minutes to completely process a patient specimen, report complete results and become ready for the next specimen.

Reagent slides   Available in a selection of 21 blood chemistries, three prepackaged panels (General Health Profile, Preanesthetic Panel and Equine Health Profile) and a Urine Protein:Creatinine Ratio. Slides are supplied in boxes and are packaged individually in sealed foils.

Calibration   Performed at IDEXX. Lot-specific information is provided with each free software upgrade.

Quality Control   VetTrol quality control fluids for blood chemistries and UCRE slides. UPRO controls for UPRO slides.

Technology   Proven dry-slide technology

Power Supply   Universal power supply

Interface Ports   RS232 port for connection to the IDEXX VetLab® System and practice-management systems

Operating Temperature   19°–27°C (66°–81°F) and relative humidity between 30% and 85%

Software   Specified to incorporate selectable Dutch, English, French, German, Italian, Japanese and Spanish interfaces.

Dimensions   Width: 18.4 inches (465 mm)
Depth: 14.3 inches (360 mm)
Height: 7.9 inches (200 mm)
Height including pipettor: 9.7 inches (245 mm)

Net Weight   31 lb (14 kg)

Accessories   Color printer, Keyboard XT 1.0 (old), Keyboard AT 2.0 (new), Keyboard drawer, VetTest® pipette, VetCom assembly, Slides

View the IDEXX VetLab® System accessories.
 
© 2008 IDEXX Laboratories, Inc.
All Rights Reserved.